4.7 Article

Coulometric microflow titrations with chemiluminescent and amperometric equivalence point detection - Bromimetric titration of low concentrations of hydrazine and ammonium

期刊

ANALYTICA CHIMICA ACTA
卷 409, 期 1-2, 页码 83-91

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0003-2670(99)00890-9

关键词

coulometric flow titration; coulometric backtitration; chemiluminescence detection; hydrazine; calibrationless determination

向作者/读者索取更多资源

A coulometric flow titration with electrogenerated bromine was combined with chemiluminescent and amperometric flow detection to determine low concentrations of hydrazine and ammonium. Hydrazine can be precisely titrated in the range between 2 mu M and 1.1 mM I with chemiluminescence detection of the equivalence points with relative standard deviations between 1.9 and 0.4% (n=5, alpha=0.05). 60 nM is the lowest concentration of hydrazine, which can be titrated. Despite quenching the chemiluminescence ammonia concentrations lower than 0.5 mM do not influence the titration results. Between pH I and 7 hydrazine can be selectively titrated in the presence of ammonium. The sum of hydrazine and ammonia/ammonium was titrated at pH 8.3 with amperometric detection. Both hydrazine and ammonia/ammonium can be determined without calibration. The amperometric detection of the equivalence points is to be preferred in the presence of ammonia. Hydrazine and ammonia were titrated precisely in their mixtures and in concentration ranges between 20 mu M and 0.5 mM. Micromolar concentrations of hydrazine and ammonia/ammonium were determined precisely and accurately in spiked water samples. (C) 2000 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据