4.4 Article

Improved MR imaging for patients with metallic implants

期刊

MAGNETIC RESONANCE IMAGING
卷 18, 期 3, 页码 287-295

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/S0730-725X(99)00135-6

关键词

artifacts; prostheses and implants; stainless steel; titanium

向作者/读者索取更多资源

Pediatric oncology patients with large metallic prostheses were imaged with one of two MR imaging techniques: 1) the tilted view-angle technique, 2) or a higher readout bandwidth technique. The tilted view-angle method uses an additional gradient in the slice selection direction during readout. The high bandwidth technique increases the readout bandwidth and shortens the echo time (TE). High bandwidth and short echo times were implemented in both T-1-weighted (T1W) turbo spin echo and turbo short tau inversion recovery (STIR) sequences. Both imaging techniques reduced the size of metal-induced image artifacts. The tilted view-angle method reduced the artifact to a greater degree but had inherent shortcomings. The reformatted images were blurred and shifted. The area of interest was often moved outside of the field of view, unless parameters were adjusted on the basis of a pre-scan calculation. The high readout bandwidth, short echo technique required no special preparation and reduced metal artifacts without image blurring. The combination of high-bandwidth, shorter echo turbo STIR and T1W turbo spin echo sequences with subtraction of pre- from post-contrast images allowed effective fat suppression without local field inhomogeneity affects. This greatly improved our ability to evaluate suspected disease near metallic implants in pediatric cancer patients. (C) 2000 Elsevier Science Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据