4.7 Article

Multivariate optimization of photochemical vapor generation for direct determination of arsenic in seawater by inductively coupled plasma mass spectrometry

期刊

ANALYTICA CHIMICA ACTA
卷 901, 期 -, 页码 34-40

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/j.aca.2015.10.020

关键词

UV photochemical vapor generation; As; One point gravimetric standard addition and internal standardization; Inductively coupled plasma mass spectrometry

资金

  1. National Natural Science Foundation of China [21205007, 21128006]
  2. China Scholarship Council [5044]
  3. Scientific Research Fund of Science and Technology department of Sichuan province [2014JY0155]
  4. Chengdu University of Technology [KYGG2014]

向作者/读者索取更多资源

Photochemical vapor generation (PVG) sample introduction coupled to inductively coupled plasma mass spectrometry (ICPMS) is described for the determination of As in seawater. A Plackett-Burman design (PBD) and central composite design (CCD) were employed to evaluate the significance of experimental variables relevant to the optimization of PVG-ICPMS detection. The impact of the saline matrix on the suppression of analyte signal was eliminated by use of a mixture of 20% (v/v) formic and 20% acetic acid (v/v) as the photochemical reductants. Optimized conditions yielded equivalent PVG generation efficiencies for As(III), As(V), monomethylarsonic acids (MMAs) and dimethylarsinic acids (DMAs), permitting direct and rapid determination of total arsenic in seawater without any other sample pretreatment. Quantitation was accomplished using one point gravimetric standard addition along with a spike of Se-82 internal standard to compensate for signal drift and fluctuation during analysis. The resulting method detection limit of 3 pg g(-1) (3 sigma) provided a 15-fold improvement over that obtained using direct solution nebulization, and is comparable to that for conventional chemical hydride generation (HG)-ICPMS. Accuracy was demonstrated by analysis of two Certified Reference Materials (NASS-6 and CASS-5 seawater) with satisfying results characterized by precisions of 3.5% and 3.2% RSD for CASS-5 and NASS-6, respectively. Crown Copyright (C) 2015 Published by Elsevier B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据