期刊
JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY
卷 11, 期 5, 页码 464-467出版社
SPRINGER
DOI: 10.1016/S1044-0305(00)00111-2
关键词
-
This study examines the feasibility of generating electrospray directly from the tip of two optical fibers bound together with Teflon tape. This approach does not require a capillary and syringe pump. The electrospray source is simply constructed by coiling the two optical fibers with a platinum (Pt) wire. The optical fibers extend beyond the Pt coil for approximately 1 cm. The sample solution is predeposited on the Pt coil by a micropipette. As the high voltage required for electrospray is applied to the coil, the sample solution moves along the grooves between the two optical fibers. A stable electrospray is subsequently generated at the tip of the fibers. The mass spectra of insulin, lysozyme, and ubiquitin are exactly the same as those obtained by conventional electrospray using a capillary and syringe pump. Rapid determination of the active ingredient in a tablet by this technique is demonstrated. (C) 2000 American Society for Mass Spectrometry.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐