期刊
TALANTA
卷 51, 期 6, 页码 1117-1123出版社
ELSEVIER SCIENCE BV
DOI: 10.1016/S0039-9140(00)00278-2
关键词
catalytic adsorptive stripping voltammetry; cobalt; methyl thymol blue
A very sensitive and selective catalytic adsorptive cathodic stripping procedure for trace measurements of cobalt is presented. The method is based on adsorptive accumulation of the cobalt-MTB (methyl thymol blue) complex onto a hanging mercury drop electrode, followed by reduction of the adsorbed species by voltammetric scan using differential pulse modulation. The reduction current is enhanced catalytically by nitrite. The optimum conditions for the analysis of cobalt include pH 9.0 (ammonia buffer), 2.0 mu M methyl thymol blue, 0.8 M sodium nitrite and an accumulation potential of -0.5 V (versus Ag/AgCl). The peak current is proportional to the concentration of cobalt over the entire concentration range tested (0.02-500 ng ml(-1)) with a detection limit of 0.005 ng ml(-1) for an accumulation time of 60 s. The method was applied to determination of cobalt in a mineral water sample and some analytical grade salts with satisfactory results. (C) 2000 Elsevier Science B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐