4.8 Article

On-site analysis of arsenic in groundwater using a microfabricated gold ultramicroelectrode array

期刊

ANALYTICAL CHEMISTRY
卷 72, 期 10, 页码 2222-2228

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac991185z

关键词

-

向作者/读者索取更多资源

Rapid on-site analysis of arsenic in groundwater was achieved with a small battery-powered unit in conjunction with a microfabricated gold ultramicroelectrode array (Au-UMEA). The sensor, consisting of 564 UME disks with a unique gold surface created by electron beam evaporation, was demonstrated to be highly sensitive to low-ppb As3+ using square wave anodic stripping voltammetry, The influence of the square wave frequency, pulse amplitude, and deposition potential on the arsenic peak stripping current was investigated, Varying those theoretical parameters yielded results surprisingly similar to those for the thin Hg film case. The performance of the Au-UMEA was evaluated for reproducibility and reliability. Three stability tests showed an average relative standard deviation of 2.5% for 15 consecutive runs. Limits of detection were investigated, and 0.05 ppb As3+ could be measured while maintaining a S/N of 3:1. Interference studies were performed in the presence of 50-500 ppb of Cu2+, Hg2+, and Pb2+. On-site analysis of groundwater containing arsenic was performed with a small battery-powered potentiostat. Quantification was done through standard additions, and these results were compared to the standard EPA methodology.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据