期刊
APPLIED SPECTROSCOPY
卷 54, 期 6, 页码 864-868出版社
SAGE PUBLICATIONS INC
DOI: 10.1366/0003702001950238
关键词
oxygen analyzer; ratiometric sensor; low-cost fluorimeter
A new oxygen analyzer combining a polymer-encapsulated dual-emission probe and a diode-based ratiometric fluorometer is described. The ratiometric fluorometer was configured to measure the relative fluorescence and phosphorescence intensity of the short-lived singlet and long-lived oxygen-quenchable triplet of [(dppe)Pt{S2C2(CH2CH2-N-2-pyridinium)}](BPh4), where dppe is 1,2-bis(diphenylphosphino)ethane. This luminescent dye was immobilized at 0.3% by weight in cellulose acetate/75% triethylcitrate and cast into a 0.5 mm-thick film. The dye molecule was excited with a blue light-emitting diode (LED) and the singlet and triplet emissions monitored by individual photodiodes at 570 and 680 nm, respectively. Oxygen can be accurately measured without interference from nonanalyte-induced intensity changes by monitoring the relative output of the two photodiodes. The output of the device was linear with oxygen concentration. The PO2(1/2) from the ratio-adapted Stern-Volmer plot was 9.6% oxygen (approximate to 73 torr), offering a dynamic range of 0-90% O-2. The device is stable and the oxygen measurements reproducible. Intentional photobleaching of approximate to 20% of the sensor dye had little effect upon the reproducibility of the oxygen measurements.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据