4.5 Article

Broad bandwidth frequency domain instrument for quantitative tissue optical spectroscopy

期刊

REVIEW OF SCIENTIFIC INSTRUMENTS
卷 71, 期 6, 页码 2500-2513

出版社

AMER INST PHYSICS
DOI: 10.1063/1.1150665

关键词

-

向作者/读者索取更多资源

Near-infrared (NIR) optical properties of turbid media, e.g., tissue, can be accurately quantified noninvasively using methods based on diffuse reflectance or transmittance, such as frequency domain photon migration (FDPM). Factors which govern the accuracy and sensitivity of FDPM-measured optical properties include instrument performance, the light propagation model, and fitting algorithms used to calculate optical properties from measured data. In this article, we characterize instrument, model, and fitting uncertaintics of an FDPM system designed for clinical use and investigate how each of these factors affects the quantification of NIR absorption (mu(a)) and reduced scattering (mu(s)') parameters in tissue phantoms. The instrument is based on a 500 MHz, multiwavelength platform that sweeps through 201 discrete frequencies in as little as 675 ms. Phase and amplitude of intensity modulated light launched into tissue, i.e., diffuse photon density waves (PDW), are measured with an accuracy of +/- 0.30 degrees and +/- 3.5%, while phase and amplitude precision are +/- 0.025 degrees and +/- 0.20%, respectively. At this level of instrument uncertainty, simultaneous fitting of frequency-dependent phase and amplitude nonlinear model functions derived from a photon diffusion approximation provides an accurate and robust strategy for determining optical properties from FDPM data, especially for media with high absorption. In an optical property range that is characteristic of most human tissues in the NIR (5 x 10(-3) < mu(a) < 5 x 10(-2) mm(-1), 0.5 < mu(s)' < 2 mm(-1)), we theoretically and experimentally demonstrate that the multifrequency, simultaneous-fit approach allows mu(a) and mu(s)' to be quantified with an accuracy of +/- 5% and +/- 3%, respectively. Although exceptionally high levels of precision can be obtained using this approach (< 1% of the estimated absorption and scattering values), we show that the absolute accuracy of optical property measurements is highly dependent on specific factors associated with instrument performance, model function relevance, and details of the fitting strategy used to calculate mu(a) and mu(s)'. (C) 2000 American Institute of Physics. [S0034-6748(00)00106-4].

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据