A new impedance cardiograph device for the non-invasive evaluation of cardiac output at rest and during exercise: comparison with the direct Fick method

期刊

EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY
卷 82, 期 4, 页码 313-320

出版社

SPRINGER
DOI: 10.1007/s004210000226

关键词

impedance cardiography; cardiac output; Fick principle; exercise; COPD

向作者/读者索取更多资源

The objectives of this study were to evaluate the reliability and accuracy of a new impedance cardiograph device, the Physio Flow, at rest and during a steady-state dynamic leg exercise (work intensity ranging from 10 to 50 W) performed in the supine position. We compared cardiac output determined simultaneously by two methods, the Physio Flow ((Q) over dot cPF) and the direct Fick ((Q) over dot cFick) methods. Forty patients referred for right cardiac catheterisation, 14 with sleep apnoea syndrome and 26 with chronic obstructive pulmonary disease, took part in this study. The subjects' oxygen consumption values ranged from 0.14 to 1.19 1 . min(-1) The mean difference between the two methods ((Q) over dot cFick - (Q) over dot cPF) was 0.04 1 . min(-1) at rest and 0.29 1 . min(-1) during exercise. The limits of agreement, defined as mean difference +/- 2SD, were -1.34, + 1.41 1 . min(-1) at rest and -2.34, + 2.92 1 . min(-1) during exercise. The difference between the two methods exceeded 20% in only 2.5% of the cases at rest, and 9.3% of the cases during exercise. Thoracic hyperinflation did not alter (Q) over dot cPF. We conclude that the Physio Flow provides a clinically acceptable and non-invasive evaluation of cardiac output under these conditions. This new impedance cardiograph device deserves further study using other populations and situations.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据