3.8 Article

Multielemental analysis in small amounts of environmental reference materials with inductively coupled plasma mass spectrometry

期刊

FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY
卷 367, 期 5, 页码 407-413

出版社

SPRINGER
DOI: 10.1007/s002160000378

关键词

-

向作者/读者索取更多资源

The lowest possible sample weight for performing multielemental trace element analysis on environmental and biological samples by ICP-MS has been investigated. The certified reference materials Bovine Liver NIST SRM 1577b, Human Hair NCS DC 73347 and Oriental Tobacco Leaves CTA-OTL-1 were applied at sample weights (1, 5, 20 and 50 mg aliquots, n = 10) which were significantly lower than those recommended with most recoveries in the range of 95-110%. Samples were digested in a mixture of nitric acid, hydrogen peroxide and hydrogen fluoride by closed-vessel microwave digestion. Multielemental analysis was performed with an optimized ICP-QMS method. Aqueous standard solutions were applied for external calibration with rhodium as the internal standard element. The detection limits varied between 0.02-0.38 mu g/g for Li, Na, Cr, Mn, Ni, Cu, Zn, Sr, Cd, Ba and Pb, and up to 1.92 mu g/g for Mg, Al, Ca, Fe and Ni. Digested human plasma samples were spiked with multielemental solution (0.5-10 mu g/L) to test the analytical method and the recoveries were 95-105% for most analytes. Our results show that in the case of homogeneous SRMs it is possible to use them in very low amounts (1-5 mg) for method development and quality control.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据