3.8 Article

Microcalorimeter energy-dispersive spectrometry using a low voltage scanning electron microscope

期刊

JOURNAL OF MICROSCOPY-OXFORD
卷 199, 期 -, 页码 37-44

出版社

BLACKWELL SCIENCE LTD
DOI: 10.1046/j.1365-2818.2000.00705.x

关键词

EDS; energy dispersive spectrometer; low-beam-energy microanalysis; low-voltage microanalysis; microanalysis; microcalorimeter; microcalorimeter EDS; mu cal EDS

向作者/读者索取更多资源

We describe the current performance of the prototype microcalorimeter energy-dispersive spectrometer (mu cal EDS) developed at NIST for X-ray microanalysis. We show that the low-energy mu cal EDS, designed for operation in the energy range 0.2-2 keV, offers significant advantages for low-beam-energy microanalysis. We present several examples in which the prototype mu cal EDS has been used to solve problems in low-voltage microanalysis, including the analysis of tungsten silicide (WSi2), titanium nitride (TiN) and barium titanate (BaTiO3) and the measurement of chemical shifts in Fe and C compounds.