Measurement of half-value layer in x-ray CT: A comparison of two noninvasive techniques

期刊

MEDICAL PHYSICS
卷 27, 期 8, 页码 1915-1919

出版社

AMER INST PHYSICS
DOI: 10.1118/1.1287440

关键词

computed tomography; half-value layer; quality assurance; acceptance testing

向作者/读者索取更多资源

The purpose of this paper is to determine the accuracy and reproducibility of two noninvasive methods of measuring half-value layer (HVL), ring and localization, compared with an invasive technique (suspending tube rotation). The ring method uses concentric aluminum rings about a CTDI ionization chamber at isocenter. Data were acquired using axial CT protocols (rotating x-ray tube, stationary patient table). The localization technique uses square aluminum sheets secured to the gantry shroud to filter the radiation beam, and a CTDI chamber suspended externally at isocenter. Data were acquired using localization image protocols (stationary x-ray tube, moving patient table). The invasive technique was similar to the localization technique except that the ion chamber was placed on the patient table and the tube rotation disabled using service software. Data for all techniques were collected on the same CT system. Independent data sr:ts were collected to determine reproducibility. Sensitivity to ionization chamber lateral displacement from isocenter was investigated. Measured HVLs (mm aluminum, mean+/-std, n=4) were 7.19+/-0.03 (ring); 7.17 +/-0.04 (localization); and 7.24+/-0.02 (service mode), which were not significantly different (p = 0.05). Displacing the chamber from isocenter changes the HVL, depending on the bow-tie filter, by as much as 5 mm aluminum. Aluminum filter to ion chamber distances of 25-35 cm provided accurate results. Both noninvasive techniques were accurate and reproducible at isocenter. However, the measured HVL was dependent upon the bow-tie filter and the lateral displacement of the ionization chamber with respect to isocenter. Greater than 2 cm off of isocenter, the ring technique did not provide accurate HVL measurements. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)03108-4].

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据