Rapid determination of lead and cadmium in sewage sludge samples using electrothermal atomic absorption spectrometry with slurry sample introduction

期刊

FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY
卷 367, 期 8, 页码 727-732

出版社

SPRINGER-VERLAG
DOI: 10.1007/s002160000464

关键词

-

向作者/读者索取更多资源

Lead and cadmium concentrations in sewage sludge samples are determined by suspending the ground samples in a solution containing 10% (v/v) concentrated hydrofluoric acid, 1% (v/v) concentrated nitric acid, 0.5% (m/v) dihydrogen ammonium phosphate and 0.1% (m/v) sodium hexametaphosphate. Aliquots of 20 mu L of these suspensions (4 mg/mL) are diluted to 1000 mu L with the same solution and then injected into the electrothermal atomizer. The drying stage is performed by programming a 400 degrees C temperature, a ramp time of 20 s and a hold time of 15 s on the power supply of the atomizer. No ashing step is used. Platform atomization is carried out at 1600 and 1800 degrees C for Pb and Cd, respectively. Calibration is performed using aqueous standards in the 5-75 and 0.2-5 mu g/L Pb and Cd ranges, respectively. Results obtained for three certified reference materials and four samples demonstrate the reliability of the procedures described.