3.8 Article

On-line complexation of zinc with 5-Br-PADAP and preconcentration using a knotted reactor for inductively coupled plasma atomic emission spectrometric determination in river water samples

期刊

FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY
卷 367, 期 7, 页码 653-657

出版社

SPRINGER-VERLAG
DOI: 10.1007/s002160000428

关键词

-

向作者/读者索取更多资源

An on-line zinc preconcentration and determination system implemented with inductively coupled plasma atomic emission spectrometry (ICP-AES) associated with flow injection (FI) was studied. The zinc was retained as zinc-2-(5-bromo-2-pyridylazo)-5-diethylaminophenol (Zn-(5-Br-PADAP)) complex at pH 9.2. The zinc complex was removed from the knotted reactor (KR) with 30% v/v nitric acid. An enrichment factor of 42 was obtained for the KR system with respect to ICP-AES using pneumatic nebulization. The detection limit for the preconcentration of 10 mL of aqueous solution was 0.09 mu g/L. The precision for 10 replicate determinations at the 5 mu g/L Zn level was 2.3% relative standard deviation (RSD), calculated with the peak heights obtained. The calibration graph using the preconcentration system for zinc was linear with a correlation coefficient of 0.9997 at levels near the detection limits up to at least 100 mu g/L. The method was succesfully applied to the determination of zinc in river water samples.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据