4.6 Article

Analysis of the results of the C-Fit uncemented total hip arthroplasty in young patients with hydroxyapatite or porous coating of components

期刊

JOURNAL OF ARTHROPLASTY
卷 15, 期 5, 页码 627-634

出版社

CHURCHILL LIVINGSTONE INC MEDICAL PUBLISHERS
DOI: 10.1054/arth.2000.4350

关键词

C-fit; uncemented; hip; hydroxyapatite

向作者/读者索取更多资源

Young patients at 2 separate centers received a C-Fit uncemented arthroplasty, with randomization to porous or hydroxyapatite coating. No difference in survival could be found between components that were porous or hydroxyapatite coated, with 27.5% being revised within 8 years, mainly for aseptic loosening. Acetabular augmentation screws or offset acetabular liners similarly did not affect outcome. Two of the offset liners had spun within the metal acetabular liners, with early failure. Some of the survivors showed poor clinical scores, but there was no correlation between clinical score and radiographic evidence of loosening. These results are comparable to the poorest reported survival figures in the literature for uncemented prostheses. Prosthesis coating could not be shown to affect component survival for this prosthesis.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据