4.5 Article

Heterogeneity in human cardiac troponin I standards

期刊

ANALYTICAL BIOCHEMISTRY
卷 284, 期 2, 页码 191-200

出版社

ACADEMIC PRESS INC
DOI: 10.1006/abio.2000.4710

关键词

-

向作者/读者索取更多资源

The LC-MS analysis of recombinant cardiac troponin I (cTnI) and cTnI extracted from human hearts showed a high degree of structural heterogeneity among all samples. The examined recombinant cTnI samples indicated posttranslational modifications, presumably due to their purification (i.e., 2-mercaptoethanol adducts and carbamylation) and related to their expression (i.e., an N-terminal expression tag), The extracted cTnI samples, while having a higher degree of structural heterogeneity, showed less structural variance between samples than the recombinant proteins. The LC-MS analysis of the extracted cTnI samples provided evidence of posttranslational modification by phosphorylation, acetylation, proteolytic cleavage, and intrachain disulfide bond formation.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据