期刊
REVIEW OF SCIENTIFIC INSTRUMENTS
卷 71, 期 10, 页码 3634-3639出版社
AMER INST PHYSICS
DOI: 10.1063/1.1313798
关键词
-
In this article a study of the transmission factor of the Omicron EA125 analyzer equipped with the universal lens is presented. The procedure is based on a model by Cross and Castle [J. Electron Spectrosc. Relat. Phenom. 22, 53 (1981)] and is applicable to every spectrometer which can be operated in the constant analyzer energy (CAE) and in the constant retarding ratio measuring mode. The advantage of the method is its independence on the sample and on the inelastic mean free path of the electrons. We find that the transmission factor for the CAE mode is proportional to E-kin(-1) for most measuring setups. This dependence is predicted by theory for an ideal analyzer. Deviations from this behavior are observed if the retarding ratio for a given kinetic energy is too small. The limit value of the retarding ratio for ideal behavior, i.e., an E-kin(-1) transmission factor, depends on the analyzer entrance slit aperture which has been selected. (C) 2000 American Institute of Physics. [S0034-6748(00)01811-6].
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐