4.5 Article

Radiation exposure to the spine surgeon during fluoroscopically assisted pedicle screw insertion

期刊

SPINE
卷 25, 期 20, 页码 2637-2645

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/00007632-200010150-00016

关键词

bone screws; fluoroscopy; ionization; lumbar vertebrae; occupational exposure; radiation

向作者/读者索取更多资源

Study Design. In vitro study to determine occupational radiation exposure during lumbar fluoroscopy. Objectives. To assess radiation exposure to the spine surgeon during fluoroscopically assisted thoracolumbar pedicle screw placement. Summary of Background Data. Occupational radiation exposure during a variety of fluoroscopically assisted musculoskeletal procedures has been previously evaluated. No prior study has assessed fluoroscopy-related radiation exposure to the spine surgeon. Methods. Bilateral pedicle screw placement (T11-S1) was performed in six cadavers using lateral fluoroscopic imaging. Radiation dose rates to the surgeon's neck, torso, and dominant hand were measured with dosimeter badges and thermolucent dosimeter (TLD) rings. Radiation levels were also quantified at various distances from the dorsal lumbar surface using an ion chamber radiation survey meter. Results. The mean dose rate to the neck was 8.3 mrem/min. The dose rate to the torso was greatest when the surgeon was positioned ipsilateral to the beam source (53.3 mrem/min, compared with 2.2 mrem/min on the contralateral side). The average hand dose rate was 58.2 mrem/min. A significant increase in hand dose rate was associated with placement of screws ipsilateral to the beam source (P = 0.0005) and larger specimens (P = 0.0007). Radiation levels significantly decreased as distance from the beam source and dorsal body surface increased. The greatest levels of radiation were noted on the side where the primary radiograph beam entered the cadaver. Conclusion. Fluoroscopically assisted thoracolumbar pedicle screw placement exposes the spine surgeon to significantly greater radiation levels than other, nonspinal musculoskeletal procedures that involve the use of a fluoroscope. In fact, dose rates are up to 10-12 times greater. Spine surgeons performing fluoroscopically assisted thoracolumbar procedures should monitor their annual radiation exposure. Measures to reduce radiation exposure and surgeon awareness of high-exposure body and hand positions are certainly called for.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据