Determination of pacing capture in implantable defibrillators: Benefit of evoked response detection using RV coil to can vector

期刊

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
卷 23, 期 11, 页码 1645-1650

出版社

FUTURA PUBL CO
DOI: 10.1046/j.1460-9592.2000.01645.x

关键词

evoked response; implantable cardioverter defibrillator; capture detection; threshold measurement

向作者/读者索取更多资源

Automatic detection of capture in ICDs would be useful for ensuring normal pacing function and lead integrity and may increase device longevity. Evoked response detection can be difficult due to postpace polarization. Polarization on the RV coil to can vector, however, should be absent when pacing with a true bipolar lead (pace tip to ring). Polarization on the HV coil to can vector should be low in an integrated bipolar lead due to the large surface area of the coil. Ventricular-paced responses were prospectively recorded in 20 patients during ICD implantation or replacement. Capture and loss of capture responses were noted during threshold searches with electrograms recorded between the RV coil and can. A detector wets designed to discriminate between capture and noncapture-paced responses using data from the first 11 patients and validated on the remaining 9. The detector had a sensitivity of 99.9% (detected capture on capture beats), and had a specificity of 100% (defected no capture on noncapture beats) for all lead configurations. There was no measurable polarization with true bipolar leads. in integrated bipolar leads, maximum polarization ranged from 0.0 to 16.7mV. In conclusion, paced evoked responses can be detected in ICDs using the HV coil to can vector using standard pacing waveforms. Special polarization reducing pacing waveforms are not required. These observations could be used to design ICDs with automatic pacing threshold defection.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.1
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据