4.3 Article

Development of a completely encapsulated intraocular pressure sensor

期刊

OPHTHALMIC RESEARCH
卷 32, 期 6, 页码 278-284

出版社

KARGER
DOI: 10.1159/000055626

关键词

cataract; glaucoma; intraocular lens; intraocular pressure; pressure sensor; telemetry; tonometry

向作者/读者索取更多资源

A completely encapsulated intraocular pressure (IOP) sensor equipped with telemetric signal and energy transfer is introduced integrated into a silicone disc for implantation into the eye. After implantation into enucleated pig eyes and into rabbit eyes in vivo, the IOP was recorded and compared to established techniques of IOP measurement. Pressure chamber tests showed that the sensor functioned correctly after biocompatible encapsulation in polydimethylsiloxane. In vivo and in vitro tests in rabbit and pig eyes demonstrated that the implanted system worked with the same precision as established techniques for IOP determination. The correlation between the measurements with the implanted device and pneumotonometry in several experiments was between 0.9 and 0.99. This device serves as a functioning model for the realization of a telemetric IOP sensor for integration into an artificial intraocular lens. Such a device will open new perspectives, not only in the management of glaucoma, but also in basic research for mechanisms of glaucoma. Copyright (C) 2000 S. Karger AG. Basel.