Subsidence resulting from simulated postoperative neck movements - An in vitro investigation with a new cervical fusion cage

期刊

SPINE
卷 25, 期 21, 页码 2762-2770

出版社

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/00007632-200011010-00008

关键词

cervical spine; cyclic testing; interbody fusion; subsidence

向作者/读者索取更多资源

Study Design. A biomechanical in vitro subsidence test of different cervical interbody fusion devices was performed new testing protocol that simulates physiologic. Objectives. To investigate the effect of simulated postoperative neck movements on the subsidence of the new WING cervical interbody fusion cage in comparison with two other cages and bone cement. Summary of Background Data. Cervical interbody fusion sometimes cause complications because of subsidence into the adjacent vertebrae with collapse of the intervertebral space. Complications such as cage dislocation or nonunion with instability also have been reported. To prevent such complications, the new WING cervical interbody fusion cage (Medinorm AG, Quierschied, Germany) has been developed. Its area of contact with the adjacent vertebrae is supposed to be large to resist excessive subsidence and small enough to prevent stress protection of the tissue growing in the cage. Methods, In this study, 24 human cervical spine specimens were tested after stabilization with either a WING, BAK/C, AcroMed I/F cage or bone cement. Then, in a new testing protocol, 700 pure-moment loading cycles (+/- 2 Nm) were applied in randomized directions (lateral bending, flexion-extension, and axial rotation alone or in combination with each other) to simulate the patient's neck movements during the first few postoperative days. Measurements of the subsidence depth (total height loss) in combination with flexibility tests (+/- 2.5 Nm) were performed before cyclic loading and after 50, 100, 200, 300, 500 and 700 loading cycles. Results: Cyclic loading caused subsidence in all four device groups, most distinct with BAK/C-cages (1.63 mm after MO loading cycles) followed by the new WING (0.90 mm) and the AcroMed (0.82 mm) cages. No statistically significant difference could be found among the three cage designs, However, all three cage types showed a significantly higher subsidence depth than bone cement (0.48 mm; P = 0.0.23 between each of the three cage-types and bone cement). A moderate correlation between bone mineral density and subsidence depth could be found only in the BAK/C group (r(2) = 0.495). A large subsidence depth after 700 loading cycles was associated with a large flexibility increase in the WING (r(2) = 0.786) and AcroMed groups (r(2) = 0.21), but with a small flexibility increase in the BAK/C group (r(2) = 0.58). Conclusions. Postoperative neck movements caused subsidence in all cervical interbody implant types. The new WING cage and the AcroMed cage seemed to have a better resistance against subsidence than the BAK/C cage. However, all three cage types had a significantly higher subsidence tendency than bone cement.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据