期刊
ENZYME AND MICROBIAL TECHNOLOGY
卷 27, 期 9, 页码 639-645出版社
ELSEVIER SCIENCE INC
DOI: 10.1016/S0141-0229(00)00196-4
关键词
human cytomegalovirus; quartz crystal microbalance; immobilization; self-assembled monolayers
A piezoelectric affinity sensor has been developed to detect distinctive antigens of the human cytomegalovirus. Either the specific antibodies or the antigen were immobilized on the gold electrode. To develop a rapid immunoassay, various assay formats were tested in relation with the different antigen composition. First, a direct assay was carried out immobilizing the specific antibody on the crystal surface by passive adsorption. Next, Protein A, thiol/poly L-lysine mixed self-assembled monolayers were tested as methods of gold modification. A competitive format was exploited by immobilization of the antigen onto the crystal activated by SAM and poly L-lysine. This procedure yielded a preliminary calibration curve. A linear range between 2.5 and 5 mug/ml of gB epitope in solution and a detection limit of 1 mug/ml were measured. (C) 2000 Elsevier Science Inc. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐