Predicting electrospray response from chromatographic retention time

期刊

ANALYTICAL CHEMISTRY
卷 73, 期 2, 页码 208-213

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac0006019

关键词

-

资金

  1. NIGMS NIH HHS [GM 44077] Funding Source: Medline

向作者/读者索取更多资源

The relationship between electrospray ionization response and HPLC retention time was explored. For the series of small peptides studied, higher ESI response was observed for analytes with longer reversed-phase HPLC retention times. This correlation existed for both experimentally measured retention times and those calculated from amino acid retention coefficients. This study is useful to ESI users since it provides a way to predict and understand anomalies in ESI response with a readily accessible experimental parameter (chromatographic retention time). In addition, the ability to correlate ESI response with chromatographic partitioning lends validity to the equilibrium partitioning model, which explains differences in ESI response on the basis of partitioning between the surface and interior phases of the ESI droplet.