Correction of H3+ contributions in hydrogen isotope ratio monitoring mass spectrometry

期刊

ANALYTICAL CHEMISTRY
卷 73, 期 2, 页码 192-199

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac000489e

关键词

-

向作者/读者索取更多资源

Two fundamentally different approaches, termed pointwise and peakwise, are currently used to correct hydrogen isotope ratio monitoring data for the presence of H-3(+) ion contributions. Consideration of the underlying assumptions shows that the peakwise approach is valid only for peaks with the same functional shape and only when background signals do not vary. The pointwise correction is much more versatile and can be used even when peak shapes and sizes, as well as background signals, vary significantly, It is not exact and is limited in accuracy by (1) the signal-broadening effects of electronic time constants, (2) the analog-to-digital conversion frequency, and (3) the highest frequency of the sample signal. To minimize errors for typical gas chromatographic signals, time constants of <500 ms and analog-to-digital sampling intervals of 250 ms are needed. Errors are further minimized by matching sample and standard peaks in both amplitude and D/H ratio. Using the pointwise algorithm, we demonstrate that a series of 14 homologous n-alkanes varying in concentration over a 5-fold range can be analyzed with a mean precision of 2.3 parts per thousand and no systematic errors.