4.7 Article

Cadmium ion-selective electrode based on tetrathia-12-crown-4

期刊

TALANTA
卷 53, 期 5, 页码 1065-1071

出版社

ELSEVIER
DOI: 10.1016/S0039-9140(00)00602-0

关键词

cadmium ion-selective electrode; PVC membrane; tetrathia-12-crown-4; potentiometry

向作者/读者索取更多资源

A new polyvinylchloride membrane sensor for Cd2+ ions based on tetrathia-12-crown-4 as an ionophore was prepared. The sensor exhibits a Nernstian response for cadmium ions over a wide concentration range (4.0 x 10(-7) to 1.0 x 10(-1) M) with a slope of 29 +/- 1 mV decade(-1). The limit of detection is 1.0 x 10(-7) M (0.01 ppm). It has a fast response time of < 10 s and can be used for at least 6 weeks without any divergence in potential. The electrode can be used in the pH range from 2.5 to 8.5. The proposed sensor shows fairly good discriminating ability towards Cd2+ ion in comparison with some alkali, alkaline earth, transition and heavy metal ions. It was successfully applied for the direct determination of Cd2+ in solution and, as an indicator electrode, in potentiometric titration of cadmium ions. (C) 2001 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据