4.4 Article

Dielectric resonances and B1 field inhomogeneity in UHFMRI:: computational analysis and experimental findings

期刊

MAGNETIC RESONANCE IMAGING
卷 19, 期 2, 页码 219-226

出版社

ELSEVIER SCIENCE INC
DOI: 10.1016/S0730-725X(01)00300-9

关键词

dielectric resonance; RF coil; modeling; B-1 field

向作者/读者索取更多资源

B-1 Field inhomogeneity and the relative effects of dielectric resonances are analyzed within the context of ultra high field MRI. This is accomplished by calculating the electromagnetic fields inside spherical phantoms and within a human head model in the presence and absence of an RF coil. These calculations are then compared to gradient echo and RARE images, respectively. For the spherical phantoms, plane incident wave analyses are initially presented followed by full wave finite difference time domain (FDTD) calculations. The FDTD methods are then utilized to examine the electromagnetic interactions between the TEM resonator and an anatomically detailed human head model. The results at 340 MHz reveal that dielectric resonances are most strongly excited in objects similar in size to the human head when the conducting medium has a high dielectric constant and a low conductivity. It is concluded that in clinical UFHMRI, the most important determinants of B-1 field homogeneity consist of 1) the RF coil design, 2) the interaction between the RF coil, the excitation source and the sample, and finally 3) the geometry and electrical properties of the sample. (C) 2001 Elsevier Science Inc. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据