4.4 Article Proceedings Paper

Requirements on amorphous semiconductors for medical X-ray detectors

期刊

THIN SOLID FILMS
卷 383, 期 1-2, 页码 132-136

出版社

ELSEVIER SCIENCE SA
DOI: 10.1016/S0040-6090(00)01592-3

关键词

X-ray; detector; amorphous semiconductors; medical imaging

向作者/读者索取更多资源

Solid state X-ray detectors based on large-area amorphous semiconductors, such as amorphous silicon or selenium, have been developed. The requirements for various applications in medical diagnosis determine the boundary conditions for different detectors. Key parameters are image-receptor size, spatial resolution, image frequency, signal-to-noise ratio, and long-term stability. Conventional thorax radiographs are 43 x 35 cm(2) in size. Therefore, to replace film, large detectors are required. Mammograms need to display microcalcifications of some 100 mum in detail. Fluoroscopic images are taken at a rate of 30 images/s, or even faster. Moreover, all detectors need to deliver optimum image quality at the lowest tolerable dose. The lifetime should be at least 10 years. In this paper we discuss different types of detectors that could fulfill those needs. Two fundamental concepts are compared: a scintillator coupled to a photodiode and a direct converting semiconductor, both being arranged on an amorphous silicon-switching matrix. (C) 2001 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据