4.2 Article

Validation of the Welch Allyn 'Vital Signs' oscillometric blood pressure monitor

期刊

JOURNAL OF HUMAN HYPERTENSION
卷 15, 期 3, 页码 191-195

出版社

NATURE PUBLISHING GROUP
DOI: 10.1038/sj.jhh.1001141

关键词

blood pressure; measurement; validation; oscillometric

向作者/读者索取更多资源

This study has determined the accuracy of the Welch Allyn 'Vital Signs' blood pressure (BP) monitoring device in an adult population, according to the criteria of the British Hypertension Society (British Hypertension Society, 1993). Validation was also classified by the American 'Association for the Advancement of Medical Instrumentation' (AAMI). The Welch Allyn 'Vital Signs' monitor achieved a BHS grade A for both systolic and diastolic BP. It also met the criteria for the AAMI protocol. The mean differences between trained observers using a mercury sphygmomanometer and the device were -1 +/- 5 mm Hg for systolic BP and -3 +/- 4 mm Hg for diastolic BP. The Welch Allyn 'Vital Signs' monitor can be recommended for clinical use in the adult population. This is one of the first oscillometric monitors designed for routine use in a hospital environment that has been found to provide an accurate assessment of BP by accepted validation procedures.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据