4.1 Article Proceedings Paper

Laser extraction of entrapped leads

期刊

PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY
卷 24, 期 3, 页码 329-332

出版社

WILEY
DOI: 10.1046/j.1460-9592.2001.00329.x

关键词

complications; lead entrapment; laser extraction

向作者/读者索取更多资源

Removing entrapped pacing and defibrillator leads has been greatly simplified by making use of the excimer laser extractor. Sixty-two leads were successfully removed, although where appropriate, portions of the lead remained in place in 10 cases. There were two serious complications, one tear of the superior vena cava during an ICD extraction and one tear of the atrium during an atrial lead extraction. Both were controlled successfully through emergency sternotomy, Although extracting entrapped leads has been much simplified by making use of the laser extractor, the process is not free of serious complications. Those involved in explanting leads should be aware of the outcomes and be prepared for emergent management.