期刊
JOURNAL OF CLINICAL MICROBIOLOGY
卷 39, 期 3, 页码 1008-1016出版社
AMER SOC MICROBIOLOGY
DOI: 10.1128/JCM.39.3.1008-1016.2001
关键词
-
类别
The performance of the Becton Dickinson BDProbe Tec ET System Chlamydia trachomatis and Neisseria gonorrhoeae Amplified DNA Assays (BD Biosciences, Sparks, Md.) was evaluated in a multicenter study. Specimens were collected from 2,109 men and women, with or without symptoms, attending sexually transmitted disease, family planning, and obstetrics and gynecology clinics. Both swab and urine samples were collected, and the results obtained from 4,131 specimens were compared to those from culture and the LCx nucleic acid amplification test (Abbott Industries, Abbott Park, III.), PCR and cytospin of the culture transport medium with chlamydia direct fluorescent antibody staining were used to adjudicate chlamydia culture-negative results, Sensitivity and specificity were calculated both,vith and without use of the amplification control (AC), with little apparent difference in the results, Without the AC result, sensitivity for C. trachomatis and N. gonorrhoeae were 92.8 and 96.6%, respectively, for cervical swabs and 80.5 and 84.9% for urine from women, C. trachomatis and N. gonorrhoeae sensitivities were 92.5 and 98.5%, respectively, for male urethral swabs and 93.1 and 97.9% for urine from men. This amplified DNA system for simultaneous detection of chlamydial and gonococcal infections demonstrated superior sensitivity compared to chlamydia culture and has performance characteristics comparable to those of other commercially available nucleic acid-based assays for these organisms.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐
暂无数据