4.7 Article

Stopped-flow injection liquid-liquid extraction spectrophotometric determination of palladium in airborne particulate matter and automobile catalysts

期刊

TALANTA
卷 54, 期 1, 页码 37-43

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0039-9140(00)00620-2

关键词

palladium; flow-injection; extraction; airborne particulate matter; automobile catalysts

向作者/读者索取更多资源

A stopped-flow injection liquid-liquid extraction (SF-EX-FIA) spectrophotometric method is reported for the determination of palladium(II), using the 2,2'-dipyridyl-2-pyridylhydrazone (DPPH) as a color forming reagent. The colored complex Pd(II)-DPPH was extracted in CHCl3 and the absorbance was monitored at 560 nm. An injection valve was used as a commutator in order to combine the stopped-flow technique with liquid-liquid extraction FI system. The calibration graph was linear up to 12 mg l(-1) (s(r) = 0.27%; r = 0.9999) with a detection limit of c(L) = 0.007 mg l(-1). The sampling rate was 20 injections per hour. The proposed method has been successfully applied to the determination of palladium in airborne particulate matter (APM) and in automobile exhaust gas converter catalysts. (C) 2001 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据