4.7 Article

A novel open-resonator system for precise measurement of permittivity and loss-tangent

期刊

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/19.918152

关键词

cavity-length variation technique; loss tangent; millimeter waves; open resonator; permittivity; quasioptical detection

向作者/读者索取更多资源

A simplified open resonator system is developed for the first time for the measurement of permittivity and loss tangent of low-loss materials with high accuracy. In this new system, a cavity length variation technique and a quasioptical detection technique are developed and implemented. The cavity length variation technique makes it possible to use only fixed frequency sources. The use of the quasioptical detection technique employing a lens refocusing system and an ultra sensitive liquid helium cooled detector allows us to utilize only one fixed frequency source in the system, This novel open-resonator system has a much simpler structure, it is inexpensive and provides same level of measurement accuracy compared to the system using a super-heterodyne receiver (two fixed frequency sources). Permittivity and loss tangent data of a number of fused silica specimens are measured using this simplified open resonator system and are compared to data obtained by the open-resonator system using a super-heterodyne receiver.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据