期刊
CHEMISTRY AND PHYSICS OF LIPIDS
卷 110, 期 2, 页码 151-164出版社
ELSEVIER SCI IRELAND LTD
DOI: 10.1016/S0009-3084(01)00132-3
关键词
phosphoinositides; lipid analysis; detection limits; MALDI-TOF mass spectrometry
Matrix-assisted laser desorption and ionization time-of-flight mass spectrometry (MALDI-TOF MS) has been recently established as a powerful tool for the analysis of biomoLecules. Here, MALDI-TOF MS was used for the detection of (poly-)phosphoinositides (PPI). PPI possess higher molecular weights than other phospholipids and a high phosphorylation-dependent negative charge. Both features affect the MALDI detection limits expressed as the minimum of analyte on the sample plate resulting in a signal-to-noise-ratio of S/N = 5. Using 2,5-dihydroxybenzoic acid (DHB) as matrix the detection limit for phosphatidylinositol (PI) is seven times higher than for phosphatidylcholine (PC) and further increases with increasing phosphorylation or in mixtures with other well-detectable pho spholipids. For phosphatidylinositol-tris-phosphate (PIP3) in a mixture with PC, the limit is about 20 times higher than for PI. The consequences for the experimental conditions are discussed. It is advisable to pre-separate PPI from biological lipid mixtures prior to the application of MALDI-TOF MS. (C) 2001 Elsevier Science Ireland Ltd. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐