4.2 Article Proceedings Paper

Intravascular micropump for augmented liver perfusion:: First in vivo experience

期刊

ARTIFICIAL ORGANS
卷 25, 期 5, 页码 392-394

出版社

BLACKWELL SCIENCE INC
DOI: 10.1046/j.1525-1594.2001.06772.x

关键词

rotary blood pump; portal hypertension; mechanical portal pumping

向作者/读者索取更多资源

The aim of this study was to assess the in vivo performance of a new microaxial rotary blood pump developed for long-term intraportal implantation. The pump, measuring 7 mm in diameter, has a single stage impeller and is powered by a microelectric motor. The pump was implanted into the portal vein in 13 large white pigs under general anesthesia. All animals recovered after the portal pump implantation. and they were observed until the pump failed. The 2 longest running pumps performed for 40 and 36 h, respectively. Either thrombus formation or technical problems, especially in the bearings, were the main causes of pump failure during the experiment. No local or systemic adverse effects were observed during the portal pumping period. Full recovery of the animals following intraportal pump implantation was achieved. However, further technical improvements to the pump are required to maintain a longer performance in vivo.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.2
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据