4.4 Article

Validation of a computerized metabolic measurement system (Oxycon-Pro®) during low and high intensity exercise

期刊

INTERNATIONAL JOURNAL OF SPORTS MEDICINE
卷 22, 期 4, 页码 291-294

出版社

GEORG THIEME VERLAG KG
DOI: 10.1055/s-2001-14342

关键词

Oxycon-Pro (R); validation; ventilation; oxygen consumption

向作者/读者索取更多资源

The aim of the study was to validate a new, fast metabolic measurement system (the Oxycon-Pro (R)) during low and high exercise intensities against the Douglas bag method. For this purpose twelve highly trained subjects performed an incremental cycle ergometer test. In the third minute of each exercise step simultaneous measurements of the Douglas bag and the Oxycon-Pro (R) were performed. No significant differences between the Oxycon-Pro (R) and the Douglas bag measurements for minute ventilation (V-E), oxygen uptake (VO2) and carbon dioxide expiration (VCO2) were found. Bland and Altman analysis of validity demonstrated minimal bias and low standard deviations. In conclusion, the results show that the Oxycon-Pro (R) is a valid apparatus for determination of minute ventilation (V-2), oxygen uptake (VO2) and carbon dioxide expiration (VCO2) during low as well as during maximal exercise intensities. Unlike with the Douglas bag method, this fast metabolic measurement system can be used for accurate and quick determination of ventilatory variables during exercise.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.4
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据