4.5 Article

Error analysis of a natural breathing calibration method for respiratory inductive plethysmography

期刊

出版社

PETER PEREGRINUS LTD
DOI: 10.1007/BF02345285

关键词

non-invasive; respiratory monitoring; respiratory inductive plethysmography; calibration

向作者/读者索取更多资源

Respiratory volumes are measured non-invasively by recording rib cage and abdominal motions using respiratory inductive plethysmography (RIP). Qualitative diagnostic calibration (ODC) of RIP is based on the natural variability in the relative rib-cage-to-abdomen contribution during tidal breathing. ODC does not require subject cooperation but it has previously been shown that accuracy may deteriorate when breathing pattern changes. The aim of this study was to investigate the causes and situations where ODC accuracy deteriorates. The ODC method was compared to PRA (calibration during voluntarily preferential rib cage or abdomen breathing) in ten adults. A reference RIP calibration was obtained from all validation data (REF). The PRA method had better accuracy than the ODC method (p < 0.01). The volumetric error ranged between 10% and 136% with ODC and between 5% and 33% with PRA, The PRA calibration factors were within 6% of those from REF, while the ODC rib-cage factor was underestimated by 15% and the abdominal factor was overestimated by 38%. Small natural variability in the relative rib-cage-to-abdomen contribution was related to poor accuracy. Each compartment's variability depended on its magnitude, which is a violation of the ODC assumptions.

作者