3.8 Article

High-performance liquid chromatographic assay of zonisamide in human plasma using a non-porous silica column

期刊

JOURNAL OF CHROMATOGRAPHY B
卷 755, 期 1-2, 页码 337-341

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0378-4347(01)00062-7

关键词

zonisamide; non-porous silica column

向作者/读者索取更多资源

A new method for measuring zonisamide (ZNS) in plasma by high-performance liquid chromatography was developed by using a 2-mum reversed-phase non-porous silica column. ZNS in plasma was first purified with a column extraction technique and injected onto the non-porous silica column. Calibration curve was linear over the concentration range of 1-80 mug/ml in plasma. The recoveries of ZNS added to plasma were more than 95.4% with the coefficient of variation less than 9.0%. We developed a rapid routine method using the non-porous silica column that was accurate and improved solvent consumption in the measurement of ZNS. (C) 2001 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据