期刊
ANALYTICAL BIOCHEMISTRY
卷 292, 期 2, 页码 216-221出版社
ACADEMIC PRESS INC
DOI: 10.1006/abio.2001.5060
关键词
-
We describe an assay scheme for glucose based on fluorescence resonance energy transfer (FRET) between concanavalin A (con A), labeled with the near-infrared fluorescent protein allophycocyanin (APC) as donor, and dextran labeled with malachite green (MG) as acceptor. Glucose competitively displaces dextran-MG and leads to reduction in FRET, assessed by time-domain fluorescence lifetime measurements using time-correlated single-photon counting. The assay is operative in the glucose concentration range 2.5-30 mM, and therefore suitable for use in monitoring diabetes control. Albumin and serum inhibit FRET but the interference can be prevented by removal of high molecular weight substances by membrane filters. APC shows promise for incorporation in an implanted glucose sensor which can be interrogated from outside the body. (C) 2001 Academic Press.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐