3.8 Article

Cylindrical diameters method far calibrating section thickness in serial electron microscopy

期刊

JOURNAL OF MICROSCOPY-OXFORD
卷 202, 期 -, 页码 468-472

出版社

BLACKWELL SCIENCE LTD
DOI: 10.1046/j.1365-2818.2001.00926.x

关键词

alignment; mitochondria; serial sections; stereology three-dimensional reconstruction; ultrathin sections

资金

  1. NIMH NIH HHS [MH/DA 57351] Funding Source: Medline
  2. NINDS NIH HHS [NS33574, NS21184] Funding Source: Medline

向作者/读者索取更多资源

An estimate of section thickness is required for measuring structures in serial section microscopy, Mean section thickness is estimated reliably by averaging the ratios of the diameters of cylindrical objects, such as mitochondria, to the number of sections they span. This cylindrical diameters method improves the accuracy of section thickness as inferred from the colour of sections floating in water. The cylindrical diameters method gives the same answer as that obtained by the minimal folds method. It is preferable because it can be done in a series that has no folds that can distort and obscure the objects that are being measured.