4.6 Article Proceedings Paper

Finite-element analysis of aortic valve-sparing: Influence of graft shape and stiffness

期刊

IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING
卷 48, 期 6, 页码 647-659

出版社

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/10.923783

关键词

aortic root; aortic valve; finite element; valve sparing; vascular graft

向作者/读者索取更多资源

Aortic valve incompetence due to aortic root dilation may be surgically corrected by resuspension of the native valve within a vascular graft. This study was designed to examine the effect of graft shape and material properties on aortic valve function, using a three-dimensional finite-element model of the human aortic valve and root. First, the normal root elements in the model were replaced with graft elements, in either a cylindrical or a pseudosinus shape. Next, the elements were assigned the material properties of either polyethylene terephthalate, expanded polytetrafluoroethylene, or polyurethane. Diastolic pressures were applied, and stresses, strains, and coaptation were recorded for the valve, root, and graft. Regarding shape, the cylindrical graft models increased the valve stresses by up to 173%, whereas the root-shaped graft model increased valve stresses by up to 40% as compared to normal. Regarding material properties, the polyurethane models demonstrated valve stress, strain, and coaptation values closest to normal, for either root shape. Graft shape had a greater effect on the simulated valve function than did the material property of the graft. Optimizing the shape and material design of the graft may result in improved longevity of the spared valve if a normal environment is restored.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据