4.2 Article

Gas transfer performance of a hollow fiber silicone membrane oxygenator:: Ex vivo study

期刊

ARTIFICIAL ORGANS
卷 25, 期 6, 页码 498-502

出版社

BLACKWELL SCIENCE INC
DOI: 10.1046/j.1525-1594.2001.06706-4.x

关键词

hollow fiber; silicone membrane; oxygenator; pediatric; cardiopulmonary bypass; extracorporeal membrane oxygenation

向作者/读者索取更多资源

Based on the results of in vitro studies of many experimental models, a silicone hollow fiber membrane oxygenator for pediatric cardiopulmonary bypass (CPB) and extracorporeal membrane oxygenation (ECMO) was developed using an ultrathin silicone hollow fiber with a 300 mum outer diameter and a wall thickness of 50 mum. In this study, we evaluated the gas transfer performance of this oxygenator simulating pediatric CPB and ECMO conditions. Two ex vivo studies in a pediatric CPB condition for 6 h and 5 ex vivo studies in an ECMO condition for 1 week were performed with venoarterial bypass using healthy calves. At a blood flow rate of 2 L/min and V/Q = 4 (V = gas flow rate, Q = blood flow rate) (pediatric CPB condition), the O-2 and CO2 gas transfer rates were maintained at 97.44 +/- 8.88 (mean +/- SD) and 43.59 +/- 15.75 ml/min/m(2), respectively. At a blood flow rate of 1 L/min and V/Q = 4 (ECMO condition), the O-2 and CO2 gas transfer rates were maintained at 56.15 +/- 8.49 and 42.47 +/- 9.22 ml/min/m(2), respectively. These data suggest that this preclinical silicone membrane hollow fiber oxygenator may be acceptable for both pediatric CPB and long-term ECMO use.