On-line bioprocess monitoring with a multi-wavelength fluorescence sensor using multivariate calibration

期刊

JOURNAL OF BIOTECHNOLOGY
卷 88, 期 1, 页码 47-57

出版社

ELSEVIER SCIENCE BV
DOI: 10.1016/S0168-1656(01)00257-7

关键词

on-line multi-wavelength fluorescence sensor; bioprocess monitoring; partial least square regression models; Pseudomonas fluorescens; in-direct measurements

向作者/读者索取更多资源

Cultivations of Pseudomonas fluorescens were monitored with a multi-wavelength on-line fluorescence sensor. The multi-wavelength fluorometer used excitation light from 270 to 550 nm with 20 nm steps and measured fluorescence emission from 310 to 590 nm. The fluorescence, on-line exhaust gas measurements and off-line analysis of nitrate, succinate, optical density and protein were compared chemometrically by multivariate calibration, i.e. computing partial least square (PLS) regression models. Based on the multivariate regression models, it was possible to determine CO2 and O-2 composition in the exhaust gas (the correlation coefficients, R-2 between the predicted values by the PLS model and the measured values was 0.97 for CO2 and 0.97 for O-2, respectively). Also to make quantitative determinations of succinate (R-2 = 0.97), protein (R-2 = 0.94), optical density (R-2 = 1.0) and nitrate (R-2 = 0.98) in the medium based on the fluorescence spectra. Only a limited data set was available but the results indicated that the sensor could indirectly determine non-fluorescent compounds, i.e. nitrate and succinate, which probably is due to the stoichiometric relationship between fluorescent cellular components and non-fluorescent compounds. Consequently multi-wavelength fluorescence is an interesting technique for a wide range of applications. (C) 2001 Elsevier Science B.V. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.5
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据