4.6 Article

A novel biosensor regulated by the rotator of F0F1-ATPase to detect deoxynivalenol rapidly

期刊

出版社

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.bbrc.2012.05.119

关键词

F0F1-ATPase; Immuno-rotary biosensor; epsilon-Subunit; Activity regulation; DON

资金

  1. Special Fund for Agro-scientific Research in the Public Interest [201103007]
  2. National Science and Technology Supported Program for the Control of Food Biological Toxin Pollution and Early Warning Technology of the Radioactive Pollution [2012BAK17B13]

向作者/读者索取更多资源

A novel biosensor (immuno-rotary biosensor) was developed by conjugating deoxynivalenol (DON) monoclonal antibodies with the rotator epsilon-subunit of F0F1-ATPase within chromatophores with an epsilon-subunit monoclonal antibody-biotin-avidin-biotin linker to capture DON residues. The conjugation conditions were then optimized. The capture of DON was based on the antibody-antigen reaction and it is indicated by the change in ATP synthetic activity of F0F1-ATPase, which is measured via chemiluminescence using the luciferin-luciferase system with a computerized microplate luminometer analyzer. 10(-7) mg/ml of DON can be detected. The whole detection process requires only about 20 min. This method has promising applications in the detection of small molecular compounds because of its rapidity, simplicity, and sensitivity. (C) 2012 Elsevier Inc. All rights reserved.