4.7 Article

Practical common-path heterodyne surface profiling interferometer with automatic focusing

期刊

OPTICS AND LASER TECHNOLOGY
卷 33, 期 4, 页码 259-265

出版社

ELSEVIER SCI LTD
DOI: 10.1016/S0030-3992(01)00035-4

关键词

common-path heterodyne interferometer; automatic focusing; phase measurement

向作者/读者索取更多资源

In modern semiconductor and optics industries, there is a strong demand for a highly sensitive and non-contact surface profilometer. This paper describes a practical heterodyne surface profiling interferometer for on-line non-contact measurement which has been developed recently. The essential feature of the profilometer is a newly designed common-path configuration to minimize the effects caused by vibration, air turbulence and other environmental variations. A single-mode frequency-stabilized laser diode (780 nm) serves as the light source to make the whole system compact (total volume 250L x 200W x 100H mm). A powerful signal processing scheme is also developed, which includes three parts: automatic voltage control, phase measurement with wide range and automatic focusing control. All these make the repeatability and stability of the profiling interferometer greatly improved. The system has vertical resolution of 0.39 nm and lateral resolution of 0.73 mum. During approximately an hour, the stability is within 1.95 nm(3 sigma). (C) 2001 Elsevier Science Ltd. All rights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据