3.8 Article

Thorium and iodine memory effects in inductively-coupled plasma mass spectrometry

期刊

FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY
卷 370, 期 5, 页码 479-482

出版社

SPRINGER-VERLAG
DOI: 10.1007/s002160100837

关键词

-

向作者/读者索取更多资源

Thorium and iodine memory effects have been characterized experimentally for inductively-coupled plasma mass spectrometry by adding ammonia gas directly to the spray chamber and nebulizing aqueous ammonia sample solutions to assess analyte memory retention sites. Thorium memory effect originates from the tendency of an unidentified thorium compound to volatilize from the spray-chamber walls, and not from Th compound adsorption to nebulizer tubing. The mass spectrometer skimmer and sampler cones, ion optics, quadrupole, and other components are not responsible for the memory effect. Unlike that of thorium the iodine memory effect originates from adsorption of iodine compounds on nebulizer tubing surfaces and from volatilization of HI and I-2 from the spray-chamber walls. Addition of ammonia sample solutions or ammonia gas directly to the spray chamber eliminated the Th and I memory effects in practical analyses, and blank levels were achieved after 2 min wash-outs. Quantitative recoveries were obtained for Th and I in reference materials.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

3.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据