4.8 Article

Standardless semiquantitative analysis of metals using single-shot laser ablation inductively coupled plasma time-of-flight mass spectrometry

期刊

ANALYTICAL CHEMISTRY
卷 73, 期 13, 页码 2959-2967

出版社

AMER CHEMICAL SOC
DOI: 10.1021/ac001272n

关键词

-

向作者/读者索取更多资源

A method has been developed that allows the accurate, standardless measurement of the elemental composition of metal samples from single laser ablation (LA) pulses. This technique provides a fast, low-sample-consumption means for the characterization of samples having a range of matrixes. The method directly compares adjusted elemental signals with the total mass spectrometric signal to produce relative percent composition information. Three mathematical techniques were used to determine the accuracy and precision of single-shot LA measurement. Comparison of the techniques showed that a linear regression calculation, which plots individual elemental signals as a function of the summed signal for all elements in the sample on a point-by-point basis during a laser ablation transient proved superior. The simultaneous extraction capability of time-of-flight mass spectrometry permits the sampling of all analytes from any temporal position within the transient laser ablation pulse, thereby reducing quantitation error. A typical concentration dynamic range of 3 orders of magnitude, from 0.1 to 100%, was achieved. However, by measuring low-abundance isotopes for matrix elements, the dynamic range of the technique was extended to 4 orders of magnitude. The new technique is largely immune to sample matrix effects commonly experienced in laser ablation, By performing a complete elemental analysis from a single ablation pulse, high spatial resolution should be achieved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.8
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据