Evaluation of multielement catheter-cooled interstitial ultrasound applicators for high-temperature thermal therapy

期刊

MEDICAL PHYSICS
卷 28, 期 7, 页码 1525-1534

出版社

AMER INST PHYSICS
DOI: 10.1118/1.1381550

关键词

prostate cancer; coagulation; thermal ablation; hyperthermia; ultrasound; interstitial

资金

  1. NCI NIH HHS [CA-09215, CA-68421] Funding Source: Medline
  2. NIDDK NIH HHS [DK-51939] Funding Source: Medline

向作者/读者索取更多资源

Catheter-cooled (CC) interstitial ultrasound applicators were evaluated for their use in high-temperature coagulative thermal therapy of tissue. Studies in ex vivo beef muscle were conducted to determine the influences of applied electrical power levels (5-20 W per element), catheter flow rate (20-60 ml min(-1)), circulating water temperature (7-40 degreesC), and frequency (7-9 MHz) on temperature distribution and thermal lesion geometry. The feasibility of using multiple interstitial applicators to thermally coagulate a predetermined volume of tissue was also investigated. Results of these studies revealed that the directional shape of the thermal lesions is maintained with increasing time and power. Radial depths of the thermal lesions ranged from 10.7+/-0.7 mm after heating for 4 min with an applied power level of 5 W. to 16.2+/-1.4 mm with 20 W. The axial length of the thermal lesions is controlled tightly by the number of active transducers. A catheter flow rate of 20 to 40 ml min(-1) (52.2+/-5.5 kPa at 40 ml min(-1)) with 22 degreesC water was determined to provide sufficient cooling of the transducers for power levels used in this study. In vivo temperatures measured in the center of a 3-cm-diam peripheral implant of four applicators in pig thigh muscle reached 89.3 degreesC after 4 min of heating, with boundaries of coagulation clearly defined by applicator position and directivity. Conformability of heating in a clinically relevant model was demonstrated by inserting two directional CC applicators with a 2 cm separation within an in vivo canine prostate, and generating a thermal lesion measuring 3.8 cmx2.2 cm in cross section while directing energy away from, and protecting the rectum. Maximum measured temperatures at midgland exceeded 90 degreesC within 20 min of heating. The results of this study demonstrate the utility of single or multiple CC applicators for conformal thermal coagulation and high temperature thermal therapy, with potential for clinical applications in sites such as prostate, liver, breast, or uterus. (C) 2001 American Association of Physicists in Medicine.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据