期刊
JOURNAL OF CHROMATOGRAPHY A
卷 924, 期 1-2, 页码 197-209出版社
ELSEVIER
DOI: 10.1016/S0021-9673(01)01017-2
关键词
stationary phases, electrochromatography; mixed-mode chromatography; electroosmotic flow; pH effects; stability studies; capillary electrochromatography
Capillary electrochromatography (CEC) requires stationary phases that enable appropriate electroosmotic propel under various conditions. Analyte retention can be controlled through hydrophobic or electrostatic interaction with the packing material. The development and characterization of new strong anion-exchange materials with additional hydrophobic moieties (SAX/C-18 mixed-mode phases) is described, The synthesis was based on polymer encapsulation of porous silica. The phases were systematically characterized by means of elemental analyses, HPLC frontal analyses and CEC experiments. The studies focused on the influence of various parameters (e.g., pH, kind of buffer, capillary wall) on the electroosmotic flow (EOF). Phases with high anion-exchange capacity generated a fast and constant EOF over a wide pH range. Long-time stability of EOF and hydrophobic retention under CEC conditions were demonstrated within the course of 100 consecutive injections. The applicability of the SAX/C-18, phases in appropriate buffer systems is demonstrated for neutral, acidic and basic compounds. (C) 2001 Elsevier Science B.V. All rights reserved.
作者
我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。
推荐