4.6 Article

A comparison of cardiac output derived from the arterial pressure wave against thermodilution in cardiac surgery patients

期刊

BRITISH JOURNAL OF ANAESTHESIA
卷 87, 期 2, 页码 212-222

出版社

ELSEVIER SCI LTD
DOI: 10.1093/bja/87.2.212

关键词

heart, cardiac output; measurement techniques, thermodilution; arterial pressure; model, computer simulation

向作者/读者索取更多资源

In three clinical centres, we compared a new method for measuring cardiac output with conventional thermodilution. The new method computes beat-to-beat cardiac output from radial artery pressure by simulating a three-element model of aortic input Impedance, and includes non-linear aortic mechanical properties and a self-adapting systemic vascular resistance. We compared cardiac output by continuous model simulation (MF) with thermodilution cardiac output (TD) In 54 patients (18 female, 36 male) undergoing coronary artery bypass surgery. We made three or four conventional thermodilution estimates spread equally over the ventilatory cycle. In 490 series of measurements, thermodilution cardiac output ranged from 2.1 to 9.3, mean 5.0 litre min(-1). MF differed +0.32 (1.0) litre min(-1) on average with limits of agreement of -1.68 and +2.32 litre min(-1). Differences decreased when the first series of measurements in a patient was used to calibrate the model. In 436 remaining series, the mean difference became -0.13 (0.47) litre min(-1) with limits of agreement of -1.05 and +0.79 litre min(-1). When consecutive measurements were made, the change was greater than 0.5 litre min(-1), on 204 occasions. The direction. of change was the same with both methods in 199. The difference between the methods remained near zero during surgery suggesting that a single calibration. per patient was adequate. Aortic model simulation with radial artery pressure as input reliably monitors changes in cardiac output in cardiac surgery patients. Before calibration, the model cannot replace thermodilution, but after calibration the model method can quantitatively replace further thermodilution estimates.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.6
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据