4.7 Article

Potentiometric Cd2+-selective electrode with a detection limit in the low ppt range

期刊

ANALYTICA CHIMICA ACTA
卷 440, 期 2, 页码 71-79

出版社

ELSEVIER
DOI: 10.1016/S0003-2670(01)01052-2

关键词

Cd2+-selective electrode; ion-selective electrodes; potentiometry; low detection limits; trace level analysis

向作者/读者索取更多资源

A recently introduced model describing the response curves of polymer membrane ion-selective electrodes (ISEs) is used here to optimize a Cd2+-ISE with a view to achieve very low detection limits. In a first step, the selectivity behavior is determined for two membranes based on the ionophore, N,N,N',N'-tetradodecyl-3,6-dioxaoctanedithioamide (ETH 5435), but having different concentrations of the ionophore and ion-exchanger. Based on these data, the optimal response curves are calculated with a model that takes into account the ion fluxes induced by interfering ions. The experimental results with different ionic backgrounds correspond very well with the predicted effects. The best lower detection limit of 10(-10) M or 11 ppt Cd2+ is achieved at pH 7 with an ionic background of 10(-4) M NaNO3. (C) 2001 Elsevier Science B.V. All lights reserved.

作者

我是这篇论文的作者
点击您的名字以认领此论文并将其添加到您的个人资料中。

评论

主要评分

4.7
评分不足

次要评分

新颖性
-
重要性
-
科学严谨性
-
评价这篇论文

推荐

暂无数据
暂无数据